Evaluation of Three Handheld Lactate Meters for Measuring Hemolymph L-Lactate i...

Evaluation of Three Handheld Lactate Meters for Measuring Hemolymph L-Lactate in Decapod Crustaceans

Author(s):

Snorre Bakke,

Publisher:

Journal of Shellfish Research (33)1

Report nr:

na

Publication type:

Vitenskaplige Artikler
|

Page number:

69-76

http://www.bioone.org/doi/abs/10.2983/035.033.0109

The aim of this study was to determine the reliability and accuracy of 3 commercially available handheld lactate
meters when measuring hemolymph L-lactate in decapod crustaceans. The instruments tested were Accutrend Plus (AP), Lactate Scout (LS), and Lactate Plus (LP), and all models were tested in triplicate and compared with a commonly used enzymatic kit for L-lactate quantification (TB-kit). The Norway lobster Nephrops norwegicus and the European lobster Hommarus gammarus were used as model organisms, with emersion as an invoker of hemolymph L-lactate. Between-method comparisons were investigated by determining correlations (Pearsons r) and concordance (concordance correlation coefficient (CCC) and Bland-Altmands95% limit of agreement (95% LOA)) between handheld instruments and TB-kit. Within-instrument variation was tested by calculating the coefficient of variation (CV), and the average absolute deviation from the mean (AD). The within-instrument variation was low for all models (CV, 3.86.1%; AD, <0.3 mM), showing high reproducibility of technical replicates. A significant correlation with TB-kit was found for all handheld instruments (AP, r¼0.970; LS, r¼0.956; LP, r¼0.950). For both AP and LS, there was a moderate and significant concordance with TB-kit (AP,CCC¼0.945; LS,CCC¼0.940). The mean difference from TB-kit was 1.42mM(95% LOA, 4.501.66mM)forAPand 1.19mM(95% LOA, 3.135.51mM)for LP. Emersion experiments with the European lobster showed that AP and LS were in line with TB-kit in terms of detecting differences in L-lactate levels between groups of animals. As a result of the high frequency of error readings from LP, this instrument was not considered reliable when measuring L-lactate in decapod crustaceans. In conclusion the handheld instruments AP and LS are found to be reliable instruments when measuring hemolymph L-lactate in decapod crustaceans. Comparison of results from different methods should be avoided, and in cases when accurate absolute measurement of L-lactate is needed, methods allowing for greater accuracy and resolution should be used.